01 COA & batch testing

What the COA for niacinamide actually contains

Technical document 01 · Reviewed 2026-11-18 · Niacinamide

On this page
  1. The certificate as a test report
  2. What each figure is measured against
  3. Release limit versus batch result
  4. What moves between batches
  5. When to order an outside test
  6. What to compare between suppliers

The certificate for a niacinamide lot reports appearance, identification by IR, assay by HPLC, the pH of a 10% solution, the melting range, loss on drying, residue on ignition, nicotinic acid, the heavy-metal panel and microbiology. Each line needs a method and a release limit, not only a number. The catalog specification is 99.0–101.0% by HPLC, pH 5.5–7.5, melting 128–131 °C, loss on drying ≤0.5%, residue on ignition ≤0.1% and nicotinic acid ≤10 ppm; the batch COA reports the lot result against those limits.

In one line

The assay says how much pyridine-3-carboxamide is present; the identity lines and the nicotinic acid figure say whether it is the right one.

The certificate as a test report

A certificate becomes auditable when every row answers three questions: what was measured, which method produced the figure, and which limit the lot was released against. The table is the catalog specification, not a batch result; the batch COA carries the result for the lot.

Identity, assay and purity · niacinamide
ParameterMethodRelease limit
Appearance / identificationVisual; IRWhite crystalline powder; conforms to reference
AssayHPLC99.0–101.0%
pH (10% solution)Potentiometry5.5–7.5
Melting rangeCapillary method128–131 °C
Loss on dryingGravimetric≤0.5%
Residue on ignitionGravimetric≤0.1%
Nicotinic acidHPLC≤10 ppm
Heavy metalsICP-MSPb ≤3 · As ≤2 · Cd ≤1 · Hg ≤0.1 mg/kg
MicrobiologyPlate count and absence testsAPC ≤1,000 cfu/g; yeast and mould ≤100 cfu/g; E. coli and Salmonella absent

The assay is the headline, but it is not the identity. HPLC with an external niacinamide standard measures the amount of material that elutes at the niacinamide retention time and responds at the detection wavelength; a peak with the right retention time and the wrong functional groups is still the wrong molecule. That is why the IR identification, the pH and the melting range sit beside the assay rather than in a footnote.

What each figure is measured against

The assay is reported on a stated basis. A figure calculated on the dried substance and a figure calculated as-is are not the same number, and the certificate should say which one applies. The 99.0–101.0% window is a release range that includes method uncertainty: an HPLC result of 100.4% is a comparison with a reference standard, not a statement that the powder is 100.4% pure.

pH is measured on a 10% solution and is an identity and quality check. A pure niacinamide solution sits close to neutral; a shift toward the acid side points to nicotinic acid or residual acid from the route, and a shift toward the alkaline side points to a nicotinate salt or an alkaline residue. The melting range does a different job: 128–131 °C is the window for niacinamide, while niacin melts near 236–239 °C, so a mixed or substituted lot usually shows a low, broad or split melt. Neither test identifies the molecule on its own, and neither replaces the IR spectrum.

Loss on drying measures water, not purity. Residue on ignition measures inorganic material left after combustion, so it catches a mineral diluent or a salt that the assay may miss. Nicotinic acid is the amide's hydrolysis product and a route impurity; at a 10 ppm limit it is the line where the method's limit of quantification matters as much as the result.

What each figure proves and what it does not
FigureWhat it provesWhat it does not prove
HPLC assayAmount of material matching the standard at the niacinamide retention timeThat the peak has the niacinamide structure; the impurity profile
IR identificationThe amide and pyridine functional groups match the referenceHow much is present; the level of nicotinic acid
pH (10% solution)The acid-base character of the lotThe identity on its own; the assay
Melting rangeThe solid's melting behaviour and gross purityA trace impurity at the part-per-million level
Loss on dryingWater and volatile materialPurity or the amount of niacinamide
Nicotinic acidThe hydrolysis and route impurity at the stated limitOther related substances the method does not separate

Release limit versus batch result

The release limit is the gate; the batch result is what the laboratory measured on the lot. Both belong on the certificate, in separate columns, with the method named. A result column that repeats the specification is a copy of the gate. Ask for the raw figure, the basis, the method and the impurity's limit of quantification.

The same rule applies to “conforms”: an IR identification is only useful if the certificate names the reference spectrum or the method used for the comparison. A pH or melting result reported without the sample concentration and the heating rate is difficult to audit; the method should state both.

What moves between batches

Assay and pH move with the neutralisation and crystallisation steps. A lot that is washed poorly can carry residual acid and sit low in pH; a lot that is overdried can lose water and read high on assay if the result is calculated as-is. Nicotinic acid moves with hydrolysis during drying and storage, which is why the impurity limit is a stability marker as well as a release test. Colour can move with heat or trace iron while the assay still passes.

Particle size is a physical variable rather than a purity one. The catalog powder is a crystalline solid, and a finer mesh changes flow, dusting and blend uniformity.

When to order an outside test

Order an independent test for a new source, a first shipment into a regulated market, a disputed result or an identity that the supplier's certificate does not prove. The useful panel is targeted: HPLC assay and nicotinic acid on the same sample, IR against a reference, pH and melting range, loss on drying and residue on ignition, ICP-MS for the metals and a microbiological panel. For a suspected niacin substitution, the melting range and the IR carbonyl region are more informative than a second assay.

Ask the laboratory to report its limit of quantification for nicotinic acid. A “not detected” entry without a stated limit describes the method's reach, not the lot. Chain of custody matters as well: the sample tested should be the retained sample, and the lot code should match the certificate and the pack.

What to compare between suppliers

Put six lines side by side: assay method and basis; IR reference; pH at 10%; melting range; nicotinic acid method and limit of quantification; and the contaminant and microbiological panel. Two quotations that match on assay but differ on the other five are not describing the same powder. Add lot traceability: the lot code on the certificate, the pack and the retained sample should be identical, and the certificate should name the manufacturer, not only the trader.

The catalog specification sits on the niacinamide product page. The forms are compared on the grade page; the tests that separate niacinamide from niacin are on the identity page; the release workflow is on the quality page. The separation behind the assay is described in the HPLC method note, and the identification in the FTIR method note.

Send the specification you intend to use — grade, quantity and destination market — and we will reply with the batch COA, the technical data sheet and a quotation.