FTIR spectroscopy for mineral ingredients

Technical note · Reviewed 18 November 2026 · All methods

FTIR spectroscopy measures the absorption of infrared light by molecular vibrations, and it is used to compare a mineral powder with a reference material. For a chelate, the carboxylate bands carry the useful information: their positions and separation give evidence of coordination. It is a fast identity check that does not measure quantity.

In short

A fingerprint method that gives evidence of magnesium–carboxylate coordination without measuring how much is there.

How the measurement is made

Infrared radiation in the mid-IR region, roughly 4,000 to 400 cm⁻¹, excites vibrational modes of the molecules in the sample, and the positions and intensities of the absorption bands report on functional groups and their environment. An attenuated total reflectance accessory presses the powder against a diamond or zinc selenide crystal, so no potassium bromide pellet is needed. The result is a spectrum that can be compared with reference materials run on the same instrument.

The carboxylate group of glycine gives two diagnostic bands: an asymmetric stretch near 1,580–1,610 cm⁻¹ and a symmetric stretch near 1,400–1,420 cm⁻¹. The separation between them is used as evidence for how the carboxylate is bound, with a smaller separation pointing to bidentate coordination and a larger one to ionic or unidentate interaction. The N–H stretching region and the C–N bands add supporting information, and water of hydration broadens the spectrum around 3,400 cm⁻¹.

Identification is by comparison: the sample spectrum against a reference chelate, against glycine, and against a physical mixture of glycine and magnesium oxide. A difference spectrum and a band-by-band comparison are more informative than a library match score. The method is fast and non-destructive, which is why it is used for incoming goods screening.

What the result proves

FTIR shows which functional groups are present and, through band shifts, gives evidence that the carboxylate is interacting with magnesium rather than sitting as free zwitterionic glycine. A physical mixture tends to show the two component spectra superimposed, while a reacted chelate shows shifted and broadened carboxylate bands. That difference is the coordination evidence the method can supply.

The spectrum is also a practical identity check against a reference lot. If the incoming material matches the reference across the fingerprint region, the probability that it is the same substance is high. If a band for a different anion or a different mineral appears, the batch has changed.

In a certificate package, FTIR supplies the bonding evidence that the elemental and ligand assays cannot. It sits alongside the magnesium result, the glycine ratio and the XRD phase screen, and each answers a different question. None of them is a complete identity proof on its own.

What it cannot tell you

FTIR is not a quantitative method. Absorbance depends on contact pressure, particle size, path length and baseline correction, so band heights do not convert into a percentage without a carefully built calibration that most identity methods do not have. It cannot give an assay, a molar ratio or an oxide content.

Bands overlap, water broadens the spectrum and different hydrate forms shift the same regions, so a spectrum can be consistent with the chelate without proving that the whole lot is chelated. A blend with a small oxide fraction can produce a spectrum close to the reference, and an amorphous material gives broad bands that are hard to compare. The library match score is a similarity measure, not a statement of purity.

The method gives no stoichiometry, no crystal phase and no trace-element information. A spectrum that matches a reference chelate does not prove every magnesium ion is bound to two glycine molecules, and it does not detect lead, arsenic or microbiological contamination. Those questions need the other methods in the file.

What it proves

  • The functional groups present in the sample, from the mid-IR absorption bands.
  • Evidence that the carboxylate group is coordinated to magnesium rather than free.
  • A comparison against a reference chelate and against a physical mixture of the components.
  • A fast identity screen for incoming goods, before the full panel is run.

What it does not prove

  • Any quantitative result: assay, molar ratio or oxide content.
  • The fraction of magnesium that is actually chelated.
  • Crystal phase, stoichiometry or trace contaminants.
  • A blend whose spectrum overlaps the reference in the fingerprint region.

What to ask the laboratory

  1. Was the spectrum acquired by ATR, and what crystal and resolution were used?
  2. Which carboxylate bands are reported, and what separation is expected for this chelate?
  3. Was the sample compared with a reference chelate, with glycine and with a physical mixture?
  4. Is the spectrum used only for identity, or is any quantification claimed?

Read the spectrum as evidence, not as a result. The useful comparison is band by band against a reference chelate and against the physical mixture, with the carboxylate separation reported in wave numbers. A library match score of 0.95 is a similarity measure and says nothing about purity. Ask which accessory was used, how the baseline was corrected and whether the reference was run on the same instrument.

Where the method is used

The method pages sit behind the COA page. The terms used here are defined in the buyer glossary, and the regulatory limits are on the compliance page.

Send the certificate or specification you are reviewing and we will name the method behind each figure.

Request a specification review

Next step

A number without a method is not a result.

Send the certificate and we will name the method, the limit and the gap.