06 Identity & adulteration

How we prove a glycine lot is glycine

Technical document 06 · Reviewed 2026-11-18 · Glycine

On this page
  1. The four layers of identity
  2. Titration and HPLC: reading the two numbers as a ratio
  3. IR, XRD and the crystal form
  4. The adulterants that fit the assay
  5. How to write the identity block
  6. Five questions for the supplier

Identity is proven in four layers: the document set, the IR spectrum, the assay by titration cross-checked by HPLC, and the impurity fingerprint of chloride, residue on ignition, heavy metals and microbiology. No single result proves glycine; the layers have to agree. The fourth layer is where a different amino acid, a salt or a glycine derivative usually shows up.

In one line

IR names the molecule, titration and HPLC measure it, and chloride and residue on ignition show what else is in the drum.

The four layers of identity

What each identity layer proves and misses
LayerTestWhat it provesWhat it misses
DocumentSpecification, COA, lot codeThe lot is described and traceableWhether the drum matches the paper
SpectrumIR against a referenceThe functional groups match glycinePurity and the level of a similar amino acid
AssayTitration, with HPLC confirmatoryThe acid–base content and the separationA non-titratable diluent or a close analogue
ImpurityChloride, residue on ignition, metals, microWhat else is in the drumThe identity of an organic analogue at low level

The layers are not interchangeable: a correct IR spectrum does not prove the assay, and a titration does not prove the spectrum. The certificate is useful when it carries all four; the buyer's own identity test is the fifth check that the paper belongs to the drum.

Titration and HPLC: reading the two numbers as a ratio

The release assay titrates the amino group on the dried basis. HPLC separates glycine from other amino acids and derivatives. Compare the two on the same basis; a dried-basis titration and an as-is HPLC result cannot form a meaningful ratio.

When the two agree within the methods' uncertainty, the lot behaves like a single amino acid. A persistent gap is a finding, not a rounding error. A high titration with low HPLC can mean a titratable impurity the HPLC method does not see, a derivatisation problem, or a different amino acid with a similar equivalent weight. A low titration with high HPLC can mean moisture or a non-titratable diluent. The next step is the chromatogram and the sample preparation, not a new assay on the same sample.

Glycine has no strong UV chromophore, so the HPLC method usually relies on derivatisation or on a detector such as refractive index, evaporative light scattering or low-wavelength UV. The certificate should name the column, the detector and the derivatisation, because a glycine result from an unnamed HPLC method is hard to audit.

IR, XRD and the crystal form

IR is the identity test on the certificate. The spectrum shows the carboxylate and ammonium bands and is compared with a reference run under the same conditions. A potassium bromide disc, an attenuated total reflectance accessory and a mull can give different relative intensities, so the certificate should state the technique and the reference.

XRD answers a different question. Glycine has more than one crystal form; alpha is the usual commercial form, and beta and gamma are known. A standard COA does not carry XRD, and a routine lot does not need it. Reach for XRD when a supplier claims a specific form, when a milled or granulated grade behaves differently from the reference, or when a dispute turns on the solid state. It is not an impurity screen, and a clean pattern does not replace the chromatogram.

If the grade is milled or granulated, ask whether the process changes the crystal form and whether the supplier has XRD data for the finished grade. A particle-size change is not a form change by itself, but the question is worth asking before a specification is fixed.

The adulterants that fit the assay

Adulterants and the lines that catch them
AdulterantWhy it can pass a rough checkWhat catches it
Other amino acids (alanine, serine)Same acid–base behaviour; similar massHPLC separation; IR differences
Glycine hydrochlorideContains glycine; the chloride is easy to missChloride; pH; IR
Sodium glycinateContains glycine; the sodium is easy to missResidue on ignition; pH
Glycine anhydrideGlycine-derived; different molecular formHPLC retention; IR
Salt diluents (NaCl, NH4Cl, sulfate)May not consume titrant; add massChloride; residue on ignition; assay balance

Alanine is the closest relative: one more carbon and the same acid–base groups, so titration alone will not separate it. Serine and valine also matter in a poorly resolved HPLC method. The chromatogram, not the percentage, is the evidence.

Glycine derivatives are another route to a misleading number. Glycine hydrochloride and sodium glycinate contain glycine and change the chloride or ash; glycine anhydride is a cyclic dimer that can form on excessive heat and has a different retention time. Each leaves a marker in the impurity panel. Salt diluents are caught by chloride and residue on ignition; a salt that lifts the ash while the assay stays in range is a specification failure.

Water is the simplest diluent. Because the assay is reported on the dried basis, a wet lot can still pass 98.5% while the as-is powder contains less glycine. Read LOD with the assay, not after it.

How to write the identity block

  • “Glycine, CAS 56-40-6, C2H5NO2, white crystalline powder.”
  • “Identification: IR spectrum conforms to the reference, technique stated.”
  • “Assay: 98.5–101.5% by titration on the dried basis; HPLC confirmatory result reported with column and detector.”
  • “Loss on drying ≤0.2%; residue on ignition ≤0.1%; chloride ≤0.1%.”
  • “Heavy metals: Pb ≤3 · As ≤2 · Cd ≤1 · Hg ≤0.1 mg/kg; microbiology as specified.”
  • “Particle size: ≥95% through 40 mesh, or the agreed granulated distribution.”
  • “Composition statement: no added other amino acids, salts or glycine derivatives.”

The composition statement turns the identity question into a specification. Sign it with the specification, not after a dispute. If the material is a blend or granulated, name the granulating agent and its amount.

Five questions for the supplier

  1. Which IR reference and technique were used, and can you send the spectrum?
  2. Which assay is the release method, on what basis, and what does the HPLC report show?
  3. What are the chloride and residue-on-ignition results for this lot?
  4. Is the material a single amino acid, or does it contain another amino acid, a salt or a glycine derivative?
  5. Which crystal form is supplied, and is XRD data available for the milled or granulated grade?

Five answers should agree with the certificate and with each other. The product page carries the catalog specification; the COA page shows where each test sits; the grade page covers the physical forms; the FTIR note explains the spectrum; the retained-sample note covers the sample chain.