01 COA & batch testing

Reading a calcium pantothenate COA: assay, calcium and rotation

Technical document 01 · Reviewed 2026-11-19 · Vitamin B5

On this page
  1. The rows that make the report auditable
  2. Dried basis and the two conversion directions
  3. Where the limit ends and the result begins
  4. What changes from lot to lot
  5. When to use an outside laboratory
  6. Six lines for a supplier comparison

A calcium D-pantothenate COA is a lot-level test report, not a restatement of the catalog. It should show the IR and HPLC identity, the HPLC assay calculated on the dried material, the calcium titration, the specific rotation, moisture, the elemental and microbiological screens, and the 40 mesh result. Each item needs a method and a limit, and the result has to be a number rather than the limit copied across.

In one line

The assay and the rotation are reported on the dried basis, and the calcium figure has to agree with them before the lot is released.

The rows that make the report auditable

Three fields per row are enough to audit a certificate: the test, the technique, and the acceptance limit. The result sits in a fourth column. A certificate that prints the specification range in the result column has told the reader what was required, not what was found.

For this material, the identity rows do more work than the assay. IR and HPLC show the pantothenate molecule; calcium and rotation show that the molecule is present as the D-calcium salt. A report that names only “vitamin B5” leaves both questions open.

Routine tests on a calcium D-pantothenate certificate
TestTechniqueLimit
AppearanceVisualWhite powder
IdentificationIR and HPLC against the referenceMatch
Assay, dried basisHPLC98.0–101.0%
CalciumComplexometric titration8.0–8.6%
Specific rotationPolarimetry after drying+25.0° to +28.5°
MoistureLoss on drying, gravimetric≤5.0%
Elemental screenICP-MSPb 3 · As 2 · Cd 1 · Hg 0.1 mg/kg
Micro screenPlate count and absence testsAPC ≤1,000 cfu/g; yeast and mould ≤100 cfu/g; E. coli and Salmonella absent
Particle sizeSieve analysis≥95% through 40 mesh

Dried basis and the two conversion directions

The assay is a dried-basis figure. The bench result is divided by the dry matter, so the moisture result is part of the assay, not an accessory. If the powder reads 95.0% as-is and loses 3.0% on drying, the dried figure is 97.9%. A 98.0% dried result at 4.0% moisture is 94.1% of the powder the production line weighs.

Rotation is measured on the dried material too. The calcium result should say whether it is as-is or dried; without that word the number cannot be compared with the assay. The ratio of calcium to assay is a quick consistency check: 100% assay corresponds to 8.41% calcium, so the ratio sits near 0.084. A certificate showing 8.6% calcium beside 98.0% assay has two numbers that do not belong to the same pure salt.

That check is not a replacement for the assay. It is a way to catch a transcription error, a blend or a different salt before the lot is used. Ask for the raw values and the calculation, not only the pass statement.

Where the limit ends and the result begins

The acceptance limit is the boundary the lot had to meet. The result is the value the laboratory obtained. A report that uses the limit as the result has collapsed the two columns and removed the evidence. The reader should be able to see both and reach the release decision without guessing.

What a lot-specific result should look like
TestLimitWhat a result looks like
Assay98.0–101.0% driedA percentage plus the HPLC method and the loss on drying
Calcium8.0–8.6%A percentage, the titration method and the basis
Rotation+25.0° to +28.5° dryThe measured angle, temperature and cell path
Moisture≤5.0%A percentage from the gravimetric test
Elemental screenEach elementA separate mg/kg figure for Pb, As, Cd and Hg
Micro screenCount and absenceThe colony count plus the absence statement

“Conforms” has its place for identity, where the result is a comparison with a reference. It has no place in the assay, calcium or moisture rows, where a number exists. For the elemental screen, a single total hides which metal was high. For micro, the count and the absence statement are two results and both belong on the page.

What changes from lot to lot

Crystallisation moves the assay, the rotation and the calcium figure together; the dryer moves the moisture result; the mill and the sieve move the 40 mesh figure. A new sieve can change that last line while the chemistry is unchanged. Bulk density and flow are not in the catalog; if the process needs them, they become additional agreed lines.

Colour is a poor test. White to off-white is broad enough to cover a normal crystallisation batch, and a cream tint can still pass. A noticeably damp or dark sample should trigger a moisture test, not a rejection based on a photograph. Trends across certificates say more than one result near an edge: falling assay, rising moisture, or rotation sitting at the bottom of the window.

The lot code is the link between the paper and the pack. It should appear on the certificate, the pouch or drum, and the retained sample. If the three do not agree, the certificate belongs to a different lot.

When to use an outside laboratory

An independent test earns its place for a new source, a first shipment into a regulated market, a disputed form, or a claim the supplier's certificate does not support. The panel should include IR, HPLC assay, calcium by ICP-OES or titration, specific rotation, loss on drying, the elemental screen and microbiology. Add a chiral method when the rotation is low or the material may be racemic.

Residual solvents are worth adding when the route uses an alcohol or another solvent that the catalog does not cover. Ask each laboratory for its detection limit; a “not detected” line without a limit describes the method, not the sample. Keep the tested sample and the retained sample from the same lot, and name the lot on the report.

Six lines for a supplier comparison

Compare form and CAS number; assay and basis; calcium; specific rotation; loss on drying; elemental, microbiological and 40 mesh results. Two certificates that agree on assay but differ on rotation or calcium are not describing the same powder.

Then look at traceability: who made the lot, which route and drying step were used, and which mill and sieve produced the powder. The catalog values are on the vitamin B5 product page. The form tests are explained on the identity page; market documents are on the compliance page; the assay technique is covered in the methods library; the release sequence is on the quality page.

Send the specification you intend to use — grade, quantity and destination market — and we will reply with the batch COA, the technical data sheet and a quotation.

Request a quotation