The certificate for an L-ergothioneine lot reports appearance, identification by 1H-NMR and by HPLC, assay by HPLC, specific optical rotation, loss on drying, residue on ignition, particle size, the heavy-metal panel and microbiology. Each row should carry the method, the release limit and the result measured on that lot. The ranges on this page are the catalog specification; the batch COA is the document that confirms a particular pack. A purity figure on its own does not establish the L form.

In one line

A 99.5% assay says how much of the molecule is present; the rotation and the NMR say which molecule, and only the lot results say what was measured.

The certificate as a test report

A certificate is only useful if it can be audited line by line. Each parameter needs the measurement, the method that produced it and the limit the lot was released against. The two tables separate the rows that describe the molecule from the rows that describe the powder.

Identity, assay and moisture · L-ergothioneine
ParameterMethodRelease limit
AppearanceVisualWhite to off-white crystalline powder
Identification A1H-NMRSpectrum conforms to the reference
Identification BHPLC retentionConforms to the reference standard
AssayHPLC≥99.5%
Specific optical rotationPolarimetry, c = 1, water≥ +122°
Loss on dryingGravimetric≤0.5%
Residue on ignitionGravimetric≤0.1%
Physical, contaminant and microbiological limits
ParameterMethodRelease limit
Particle sizeSieve analysis≥95% through 40 mesh
Heavy metalsICP-MSPb ≤3 · As ≤2 · Cd ≤1 · Hg ≤0.1 mg/kg
MicrobiologyPlate count and absence testsAPC ≤1,000 cfu/g; yeast and mould ≤100 cfu/g; E. coli and Salmonella absent

What the assay and the identification actually measure

The assay is a mass-purity figure. It answers how much of the declared molecule is present, not whether that molecule is the L enantiomer. Two numbers can sit at 99.5% while one is the natural configuration and the other is a racemate: the assay does not see the difference.

The basis belongs beside the result. A figure reported as-is includes the water that loss on drying measures; a figure on the dried basis excludes it. At the 0.5% loss-on-drying ceiling the two differ by about half a percentage point — 99.5% on the dried basis is roughly 99.0% as-is. That gap is the whole distance between a passing and a failing lot when the limit is 99.5%, so the certificate has to say which basis was used.

The assay method matters as much as the number. Area normalisation cannot see water, inorganic residue or a non-chromophore impurity; an external standard reports mass purity and depends on its own purity. Ask which column, wavelength and reference standard sit behind the figure.

Identification is a separate row and should be read as one. A 1H-NMR run in D2O shows the whole proton set, including the nine-proton trimethylammonium signal that the betaine structure produces; a spectrum matched against a reference is evidence for the molecule, not for its optical purity. HPLC retention adds a second, orthogonal look at the same sample. Neither substitutes for the rotation.

Release limit versus batch result

The release limit is the acceptance boundary; the batch result is what the laboratory measured. Both belong in separate columns, with the method next to them. A result column filled with the specification range is not a result — it is a copy of the gate, and it tells the buyer nothing about the pack being shipped.

Three rows deserve the closest reading. Specific rotation should be reported with the conditions — wavelength, temperature, concentration and solvent — because the number has no meaning without them. Loss on drying is both a purity correction and the first signal of a hygroscopic solid that has taken up water in handling. Residue on ignition is small in absolute terms, at 0.1%, but it is the row that catches inorganic carry-over from a synthesis or a recovery step.

Heavy metals and microbiology are panel results, not single figures. Each element should appear against its own limit, and the microbiology section should name the tests performed rather than print one word. “Conforms” is a conclusion; the reader should be able to reach it from the numbers on the page.

What moves between batches

Assay and rotation track the isolation and the final crystallisation. Rotation is the sensitive one: a lot that passes assay while the rotation drifts toward the limit is telling you something about the isomer, not about the yield. A single low reading deserves a repeat with a fresh solution before it becomes a rejection.

Loss on drying moves with the drying step and, later, with the pack. A rising LOD across shipments that were opened in a humid room is a handling signal, not a manufacturing one. Colour is qualitative and often more revealing than a number: a tint that develops while the assay still passes usually points to oxidation or to a metal trace, and it should trigger a question rather than a quiet acceptance.

Particle size and bulk density move with milling and sieving. They matter downstream: a specification that fixes 40 mesh but says nothing about density can still deliver two powders that run differently on the same filler.

When to order an outside test

Order an independent test for a new source, a first shipment into a regulated market, an isomer claim the supplier's certificate does not prove, or a disputed result. The useful panel is built around the questions the certificate leaves open: HPLC purity with the related-substance profile, polarimetry under the declared conditions, 1H-NMR, ICP-MS for the metal panel and a moisture determination, all on one submitted sample.

If the rotation is ambiguous, chiral HPLC is the referee: it can resolve the two enantiomers and report an enantiomeric ratio that polarimetry alone only implies. Ask the laboratory to state its limit of detection for any impurity screen; a “not detected” line without a limit describes the method, not the sample.

Chain of custody matters as much as the panel: the sample tested should be the sample retained, and the lot code on the report should match the pack. Keep the retained sample until the lot is consumed.

What to compare between suppliers

Put six lines side by side: the isomer named; the assay with its method and basis; the specific rotation with its conditions; loss on drying; the related-substance profile; and the 1H-NMR identity. Two quotations that agree on the assay but diverge on the other five are not describing the same material.

The second layer is traceability. The lot code should be identical on the certificate, the pack and the retained sample, and the certificate should name the manufacturer rather than only the trader. The specification sits on the L-ergothioneine product page; the forms are compared on the grade page; the tests that separate the L form from a blend are on the identity page; the release workflow is described on the quality page; and the chromatography itself is explained in the HPLC method note.

Send the specification you intend to use — grade, quantity and destination market — and we will reply with the batch COA, the technical data sheet and a quotation.