01 COA & batch testing

What the COA for casein phosphopeptides actually contains

Technical document 01 · Reviewed 2026-11-19 · Casein Phosphopeptides

On this page
  1. The certificate as a test report
  2. The protein factor, the peptide fraction and the basis
  3. Release limit, catalog range and lot result
  4. What moves between lots
  5. When an outside test earns its fee
  6. What to compare between suppliers

A casein phosphopeptide (CPP) certificate reports protein by Kjeldahl, peptide content by HPLC, calcium by titration, ash, moisture and the contaminant panel. The first three lines are not interchangeable: protein includes the peptides, peptide measures the fraction the HPLC method counts, and calcium measures the mineral the phosphoserine clusters carry. The limits below are the catalog specification; a batch COA confirms a particular lot.

In one line

The CPP certificate turns on three figures that are not interchangeable: protein by Kjeldahl with its factor and basis, peptide content by HPLC, and calcium by titration, read beside the ash.

The certificate as a test report

A CPP certificate is read in two blocks. The composition block carries protein, peptide, calcium, ash and moisture; the contaminant block carries heavy metals, microbiology and particle size. Each line needs its method and release limit, and the measured result belongs in its own column. A certificate that prints only “conforms” asks the reader to accept the conclusion without the evidence.

Composition, contaminants and physical parameters · casein phosphopeptides, catalog limits
ParameterMethodRelease limit
AppearanceVisualOff-white to light tan powder
ProteinKjeldahl≥80.0%
Peptide contentHPLC peptide distribution≥70.0%
CalciumComplexometric titration2.0–4.0%
AshGravimetric, furnace≤10.0%
MoistureLoss on drying≤5.0%
Heavy metalsICP-MSPb ≤3 · As ≤2 · Cd ≤1 · Hg ≤0.1 mg/kg
MicrobiologyPlate count and absence testsAPC ≤1,000 cfu/g; yeast and mould ≤100 cfu/g; E. coli and Salmonella absent
Particle sizeSieve analysis≥95% through 80 mesh

The method column decides whether two certificates can be compared. A calcium figure by titration and one by ICP-OES are both valid, but they are not the same measurement; the same is true of peptide by HPLC and protein by Kjeldahl.

The protein factor, the peptide fraction and the basis

Kjeldahl measures nitrogen and multiplies it by a factor. For milk protein the factor is 6.38; the general 6.25 understates a dairy protein by about 2%. The factor must be printed beside the protein figure, because 80.0% with 6.25 and 80.0% with 6.38 are different amounts of nitrogen.

The basis is the second half of the figure. The catalog sets protein at ≥80.0% and moisture at ≤5.0%; a dry-basis result of 80.0% at the moisture ceiling corresponds to about 76.0% as-is. The certificate and the purchase order should both state the basis, because a four-point difference is enough to fail an agreed release limit.

Peptide content is a separate measurement, not a second protein figure. The HPLC distribution separates the peptide fraction from free amino acids and salts, and the result is reported as a percentage of the powder. Protein and peptide overlap: the peptides are part of the protein, so an 80% protein figure and a 70% peptide figure do not add to 150% of the material. The gap between them is where free amino acids and non-peptide solids sit.

Calcium is measured by complexometric titration and reported as the element. The ash ceiling of 10.0% is consistent with the mineral load: at 4.0% calcium the oxide alone accounts for about 5.6% of the powder, and the phosphate adds to it. Calcium, ash and phosphorus, when it is reported, belong together.

Release limit, catalog range and lot result

The catalog range is the window the grade is sold inside, the release limit is the narrower boundary agreed for a market or a customer, and the lot result is the measurement. The first two are promises; only the third is evidence.

What the result column should contain
ParameterCatalog limitResult column
Protein≥80.0%A measured percentage with the factor and the basis
Peptide≥70.0%A measured percentage with the HPLC integration range
Calcium2.0–4.0%A measured percentage and the titration method
Ash≤10.0%A measured percentage
Moisture≤5.0%A measured percentage
Heavy metalsPer elementA figure in mg/kg for each element
MicrobiologyCounts and absenceThe count and the absence statement for the lot
Particle size≥95% through 80 meshThe measured percentage through the sieve

A result column that repeats the catalog range is not a result. For microbiology the evidence is a count plus an absence statement; for heavy metals it is a figure per element, not a single total.

What moves between lots

The raw material is milk, and milk is seasonal. Feed, stage of lactation and herd management move the protein and the mineral load in the casein, and the hydrolysis and membrane steps can only work with what arrives. The hydrolysis time and the enzyme dose move the peptide distribution; a longer hydrolysis shifts the profile toward smaller peptides and can raise the peptide figure while the protein figure stays inside its window. A calcium adjustment, where it is used, moves calcium and ash together.

The dryer moves moisture, colour and flow, and milling moves the sieve result. A hotter inlet gives a drier, free-flowing powder that wets slowly; a gentler dry holds more moisture and disperses more easily. Neither is an identity failure, but they are not interchangeable in a filling line. The lot code on the certificate should match the pouch, the drum and the retained sample.

When an outside test earns its fee

Order an independent test for a new source, a first shipment into a regulated market, a peptide claim the certificate does not prove, or a disputed result. The panel should answer the open question: protein at a second laboratory with the factor stated, peptide by HPLC, calcium by titration and phosphorus by ICP-OES, ash, moisture, heavy metals, microbiology, and a melamine and cyanuric acid screen.

The melamine screen matters because Kjeldahl measures nitrogen, not protein. Melamine is about 67% nitrogen by mass, so a small addition lifts the protein figure without adding a peptide. Urea does the same on a smaller scale. Neither is caught by a protein result; both are caught by the nitrogen source and the amino acid profile. Ask the laboratory for its detection limit, because a “not detected” line without one describes the method, not the sample.

What to compare between suppliers

Put seven lines side by side: protein with its factor and basis; peptide with the HPLC method and integration range; calcium and phosphorus; ash; moisture and particle size; the allergen statement naming milk; and the traceability from the certificate to the pack. Two offers that agree on 80% protein and disagree on the peptide method or the calcium figure are not describing the same powder.

The peptide figure separates CPP from a casein powder; the amino acid profile and the calcium-phosphorus pair separate it from another hydrolysate, and the calcium figure shows whether the mineral load is part of the material or an addition. The catalog specification sits on the casein phosphopeptide product page; the tests that separate CPP from casein are on the identity page; the titration and HPLC methods are described in the methods section; the release workflow is on the quality page.

Send the specification you intend to use — grade, quantity and destination market — and we will reply with the batch COA, the technical data sheet and a quotation.

Request a quotation